28 • Here’s How! A QuickGuide for ColorQuartet
ColorQuartet 5.0 Pro
259
To white point-calibrate your flatbed scanner
NOTE: before calibrating, make sure the standard mounting plate is in the scanner, not the
optional magnetic mounting plate system or (for ScanMate F8 Plus) the optional ScanDot
mounting plate.
1 • From ColorQuartet’s Calibration menu, choose Calibrate White.
2 • Click on OK to confirm.
The scanner will automatically calibrate itself for transmission and reflec-
tion.
A message will appear to tell you whether calibration succeeded.